Effects on Unplanned Extubation, Oral Mucosa, and Facial Skin Integrity of New Method to Secure Endotracheal Tube

ÀÓ»ó°£È£¿¬±¸ 2009³â 15±Ç 3È£ p.49 ~ 59
KMID : 1004620090150030049
±èÁ¤¼÷ ( Kim Jeong-Sook ) - Àü³²´ëÇб³º´¿ø

ÀÌÀº¼÷ ( Lee Eun-Sook ) - Àü³²´ëÇб³ ÀÇ°ú´ëÇÐ °£È£Çаú
¹ÚÁøÇâ ( Park Jin-Hyang ) - Àü³²´ëÇб³º´¿ø °£È£ºÎ

Abstract

Purpose: Examination was done of the effects on unplanned extubation, oral mucosa, and facial skin integrity in patients with oral intubation secured with a new method using Multifix Endo II.

Methods: A prospective quasai-experimental design was used. Data were collected during a 3-week EICU admission period between Jun. 1 and Nov. 30. Two hundreds seventeen (conventional taping group: 105, Multifix Endo II group: 112) orally intubated adults were enrolled in the study at the EICU of the C University Hospital.

Results: The incidence of unplanned extubation (¥ö2=11.580, p=.003) and scores for the facial skin integrity impairment of the Mutifix Endo II group (t=5.28, p=.000) were significantly lower than those of the conventional taping group even though the scores for oral mucosa impairment were not different. The nurse¡¯s evaluation of the clinical effectiveness of the two methods including convenience, comfort, safety, and satisfaction were also higher for the new securement method compared to the conventional taping group.

Conclusion: The securement method using Mutifix Endo II was more effective than the conventional method in the prevention of unplanned extubation and maintenance facial skin integrity.
KeyWords
±â°ü Æ©ºê, ¹ß°ü, ÇǺΠÅëÇÕ¼º
Endotracheal tube, Extubation, Skin integrity